ASTM F3275-22 - Standard Guide for Using a Force Tester to Evaluate Performance of a Brush Part Designed to Clean the Internal Channel of a Medical Device

Standard Guide for Using a Force Tester to Evaluate Performance of a Brush Part Designed to Clean the Internal Channel of a Medical Device

Format
Availability
Price and currency
English PDF Redline
Immediate download
77.04 USD
English PDF
Immediate download
64.20 USD
English Hardcopy
In stock
64.20 USD
Standard number:F3275-22
Released:01.01.2022
Status:Active
Pages:12
Section:13.02
DESCRIPTION

1.1 Brushes used to clean a medical device after clinical use play an important role in effective reprocessing. This guide describes methods for characterizing, under prescribed laboratory conditions, the efficacy of brush parts designed to clean the internal channels of medical devices. The methods utilize a force tester to mechanically actuate a brush part within a channel: (1) Methods to measure, at an established speed, the force required to move a brush within a channel; (2) Methods utilize the same force tester and protocols to measure soil removal from a soiled tube, another indicator of performance.

1.2 Inclusions: 

1.2.1 This guide describes objective, quantifiable, and reproducible methods for evaluating the cleaning characteristics of a brush part under prescribed laboratory conditions, with test methods that simulate the cleaning challenge of a defined target area(s) of a medical device. This also makes possible the comparison of one design of a brush part to another.

1.2.2 In this guide, a brush part is one that is intended to be moved within a tube.

1.2.3 Tubes used for testing described in this guide are cylindrical and uniform in diameter. The test methods describe may not apply to non-cylindrical tubes.

1.2.4 By use of this guide, medical device manufacturers can characterize the brush part designed for cleaning their device.

1.2.5 By use of this guide, manufacturers of cleaning brushes can evaluate and characterize the cleaning performance of their brushes for the target area(s) of medical device(s), including allowing a comparison with existing brush part designs offered on the market. Further, they are able to evaluate modifications to designs and construction that might improve performance.

1.2.6 This information can also be shared with the users of the brushes (medical device reprocessors) to help them evaluate the performance of commercially available brushes.

1.3 Exclusions: 

1.3.1 This guide does not assess potential damage that may be inflicted by the brush. For instance, brushes with rigid bristles (for example, stainless steel or other metals) or other abrasive materials are more likely to damage medical devices than brushes with flexible bristles (for example, nylon) or more pliable materials. Potential damage from more abrasive materials should be assessed.

1.3.2 This guide does not specify acceptance criteria, and the results will be dependent on the specific parameters (for example, test soil, drying time, channel inside diameter and material, and so forth) that are tested.

1.3.3 This guide is not intended to constitute all steps required to conduct validation of cleaning instructions for a medical device, including the use of brushes for this purpose, but provides methods that may be part of a broader protocol to conduct a complete cleaning instructions validation.

1.3.4 If a brush is intended to clean a specific device(s), cleaning validation shall include testing with that device(s).

1.4 Units—The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.