ASTM E3263-22E01 - Standard Practice for Qualification of Visual Inspection of Pharmaceutical Manufacturing Equipment and Medical Devices for Residues

Standard Practice for Qualification of Visual Inspection of Pharmaceutical Manufacturing Equipment and Medical Devices for Residues

Format
Availability
Price and currency
English PDF
Immediate download
87.74 USD
English Hardcopy
In stock
87.74 USD
Standard number:E3263-22E01
Released:01.05.2022
Status:Active
Pages:21
Section:14.01
Keywords:attribute agreement analysis; logistic regression; maximum safe surface; residue visual inspection; visual residue limit; VRL ;
DESCRIPTION

1.1 This practice provides statistically valid procedures for determining the visual detection limit of residues and the qualification of inspectors to perform the visual inspection of pharmaceutical manufacturing equipment surfaces and medical devices for residues.

1.2 This practice applies to pharmaceuticals (including active pharmaceutical ingredients (APIs); dosage forms; and over-the-counter, veterinary, biologics, and clinical supplies) and medical devices following all manufacturing and cleaning. This practice is also applicable to other health, cosmetics, and consumer products.

1.3 This practice applies to many types of chemical residues (including APIs, intermediates, cleaning agents, processing aids, machining oils, and so forth) that could remain on manufacturing equipment surfaces or medical devices that have undergone all manufacturing steps including cleaning.

1.4 This practice applies only to equipment or devices that have been justified through a Quality Risk Management program to have an acceptable hazard analysis, have cleaning processes that are repeatable and validated and where Visual Inspection can be relied upon to determine the cleanliness of the equipment at the residue limit justified by the HBEL.

1.5 The values stated in International System of Units (SI) units are to be regarded as standard. No other units of measurement are included in this standard.

1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.