ASTM F2052-26
Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices and Other Items Intended for the B<inf>0</inf> Near Field Magnetic Resonance Environment
| Standard number: | ASTM F2052-26 |
| Released: | 01.11.2026 |
| Status: | Active |
| Pages: | 11 |
| Section: | 13.01 |
| Keywords: | medical device; metals (for surgical implants and medical devices); MRI (magnetic resonance imaging); MR safety; |
| ASTM number: | ASTM F2052 |
| Supersedes: | F2052-21 |
1.1 This test method covers the measurement of the magnetically induced displacement force produced by static magnetic field gradients (spatial field gradient) on medical devices and other items intended for MR conditional use in the B0 near field (i.e., the space around an MR equipment where the static magnetic field strength (B0) is >200 mT) of the magnetic resonance environment.
1.2 “Medical devices and other items” will be referred to as “items” for the remainder of this test method.
1.3 This test method does not address other possible safety issues which include, but are not limited to: issues of magnetically induced torque, radiofrequency (RF) heating, induced heating, acoustic noise, interaction among items, and the functionality of the item in the magnetic resonance (MR) environment.
1.4 This test method shall be carried out using a cylindrical whole-body MR equipment with a static magnetic field oriented horizontally and parallel to the MR equipment bore.
1.5 The results can be applicable to whole-body clinical MR equipment (e.g., 1.2 T, 1.5 T, 3.0 T, 7.0 T MR equipment). The results could be applicable to other types of MR equipment including head-only or extremity clinical MR equipment.
1.6 When an item is tested according to this standard in the B0 near field (ASTM F2052), there is no need for further testing in the B0 far field according to Test Method F3715.
1.7 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.8 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.9 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.